Medical Waste Sterilization

Scope of application: Medical Waste Sterilization

Processing Loads: Performs sterilization and decontamination of wastes with infection risk in hospital, laboratory and medical field

Processes: Loading, shredding, sterilization, cooling, unloading, washing

 

 

Stok kodu: RST Series Kategoriler:

Açıklama

Sterilization of Medical Waste

The process of making infected wastes harmless by exposing the infected wastes (medical wastes) to high pressure steam and temperature for a certain period of time is defined as sterilization. In other words, it is the process of inactivating microorganisms with moisture, heat and pressure.

In the disposal of medical wastes originating from health institutions; Sterilization of wastes with high heat and steam is the most effective method. The medical waste sterilization method recommended by the International Health Organization is a solution that is easy to use compared to incineration, the training of the staff is fast and the investment and operating costs are much lower.

According to the Medical Waste Control Regulation, it is obligatory to have a waste shredding mechanism as well as a sterilization autoclave in medical waste sterilization facilities. Medical waste shredders in the medical waste shredding unit are located at the end or in front of the sterilization section. If the medical waste shredding unit is used before the sterilization unit, this unit is also subjected to the sterilization process at the end of the process.

Ek bilgi

capacity

20 kg/h-500 kg/h

Cover System

Automatic cover system in vertical and horizontal direction

Warranty

two years

Product Catalog

Technicial Specifications

  • In addition to the versions varying from 160 to 10,000 liters developed by RST MEDICAL, the load capacity can be measured according to the needs of the customer.
  • The interior surface of the chamber is mechanically polished to a roughness less than Ra <0.4 μm.
  • Automatic vertical sliding or semi-automatic folding doors are available in 1 or 2-door versions, depending on the customer’s needs to optimize the available space.
  • At least 5 PT100 RTD probes (3 flexible + 2 fixed) for temperature control in chamber and product.
  • Automatic sterilization of the air filter is completed with the fixed thermo-probe in the filter bell to monitor the temperature inside.
  • 2 steps for verification sensors.
  • Door seal provided with a seal with a unique RST MEDICAL profile.
  • Seals patented by RST MEDICAL and guaranteeing 100% door tightness. These seals are FDA compliant and also USP Class VI compliant.
  • All stainless steel 316L- 1.4435 and 316L – 1.4404 hydraulic system with unique quick couplings and silicon seals.
  • Equipped with one or more fans depending on the size of the chamber, with the ventilation system, it provides a better seal in the chamber,